Channel Master Operation Manual
page 72 RD Instruments
10.9 Vertical Beam Commands
The following commands define the criteria used to set the Vertical Beam
data.
10.9.1 Standard Vertical Beam Commands
This section lists the most often used Vertical Beam commands.
VD – Vertical Beam Data Out
Purpose Selects the data types collected by the Channel Master.
Format VD abc def ghi
Range Firmware switches (see description)
Default VD 111 000 000
Recommended Setting. The default setting for this command is
recommended for most applications.
Description VD uses firmware switches to tell the Channel Master the
types of data to collect. The Channel Master always collects
header data, fixed and variable leader data, and checksum
data. Setting a bit to one tells the Channel Master to collect
that data type. The bits are described as follows:
a = amplitude d = Reserved g = Reserved
b = commands e = Reserved h = Reserved
c = status f = Reserved I = Reserved
Notes 1. Each bit can have a value of one or zero. Setting a bit to
one means output data, zero means suppress data.
2. If VP = zero, the Channel Master does not collect vertical
beam amplitude or command data.
3. Spaces in the command line are allowed.